Robotic navigation is among the tools Sanjay Kumar Khurana, whose orthopaedic practice is in Santa Monica and Marina del Rey, brings to the operating table. Put plainly, surgical navigation is a way of showing a surgeon, on a screen, where an instrument is in relation to the patient's own anatomy. A camera system follows the tool, a set of images supplies the anatomy, and software lays one over the other so the tip of the instrument can be seen against structures the eye cannot reach.
What surgical navigation means
Three parts make a navigation system work. The first is an image of the patient, usually from CT or MRI. The second is a tracker that measures where each instrument sits in space. The third is the software and display that combine the two. A 2013 engineering paper in the Journal of Lightwave Technology1 calls the near infrared optical tracker the part that largely decides how accurately surgical tools are followed, and notes that such trackers are widely used in biomedical work.
The idea is related to, but distinct from, a surgical robot. Navigation reports position; a robot moves instruments. Many modern systems do both, which is why the two words often travel together.
How an optical tracker sees an instrument
The tracker described by Rongqian Yang and colleagues uses two near infrared cameras set apart like a pair of eyes1. Each surgical tool carries markers that show up brightly in infrared light. Before use, the cameras are calibrated against a specially designed infrared calibration board, and the whole system is then calibrated from those results.
During tracking, each camera records the markers as points in its own image. A two step stereo matching method pairs the points seen by the left camera with those seen by the right, down to fractions of a pixel. From each matched pair the software computes the marker's three dimensional coordinates, and from the markers it works out where the tool is, continuously and in real time1.
- 1Calibrate Two near infrared cameras are calibrated with a purpose built board
- 2Detect Each camera records the bright markers fixed to a surgical tool
- 3Match A two step method pairs left and right image points to subpixel precision
- 4Locate The three dimensional coordinates of each marker are computed
- 5Follow The tool's position is updated continuously on the navigation display
How precise a tracker can be
Precision in a laboratory is measured by moving markers to known places and comparing. In the 2013 study, three markers were checked at 150 positions, and their average tracking errors came to a few thousandths of a millimeter1. Those figures describe how well the cameras locate a marker, which is the foundation for everything the screen shows afterward.
Images taken in the room
A tracker is only as useful as the anatomy it is matched to. Scans taken days earlier may no longer match the body once the patient lies on the table and the work has begun. In a report by Northwestern Memorial Hospital surgeons, published in Skull Base in 20112, the authors described scanning patients during skull base and sinus operations with three portable CT machines, and found that each new scan could update the image guidance system as the operation went on.
The study compared a conventional multidetector scanner with two flat panel cone beam units. The conventional machine showed soft tissue; the cone beam devices generally gave lower radiation doses and less interference from metal2. The team judged each scanner on its technical demands, ease of use, image characteristics and how well it fed the guidance system. The history behind these machines is told in X rays in surgery.
Why a scan is repeated during an operation
Surgery in narrow regions such as the skull base, the Northwestern authors wrote, depends on feedback about anatomic landmarks as the work proceeds2. A fresh scan tells the surgeon how far dissection has actually gone and resets the navigation map to match. The repeated image is a correction for change: tissue removed, structures shifted, a patient moved.
Navigation and the robotic arm
Navigation often appears alongside robot assisted surgery. CommonSpirit Health's patient guide3 describes the surgeon seated at a console, viewing a high definition three dimensional picture of the operative field, while hand, wrist and finger movements are translated into small movements of instruments on robotic arms. A team at the table assists, changes instruments and manages the system. The history of those machines is covered in surgical robots.
From screen to headset
A newer branch moves the navigation picture from a monitor to the surgeon's eyes. A 2017 New Scientist report4 described Scopis, a software firm that had built an augmented reality system to help spinal surgeons follow the position of a patient's vertebrae during an operation. At UC Davis Health, a 2024 account says, surgeons wear goggles that project CT and MRI data into view to locate instruments and implants and to show drilling and cutting guides5. The goggles turn data that once sat on flat monitors into a three dimensional reference in the surgeon's line of sight5. This line of work is described further in augmented reality in surgery.
“We can't trust these systems at the moment”Shafi Ahmed, surgeon at the Royal London Hospital, 2017, New Scientist
What tracking cannot correct
Tracking measures markers, not tissue. If the image beneath the markers is out of date or misregistered, a precise tracker will show a precise position on the wrong map, which is why the New Scientist report stressed that any overlay meant to sit exactly on a patient must be pinpoint accurate4. In the same piece the surgeon Shafi Ahmed doubted that such aids could stand in for a specialist's hands on experience.
Display is a limit too. Writing for Smithsonian magazine in 2015, Brian Handwerk quoted the Stanford imaging researcher Sandy Napel, who said experts are already good at reading flat images, so three dimensional displays may add little in some tasks6. Navigation narrows uncertainty; it leaves the surgeon still responsible for knowing the anatomy, for judging when the map and the patient disagree, and for deciding when to stop and scan again.
Sources cited
- Design of an Accurate Near Infrared Optical Tracking System in Surgical Navigation. Optica, 2013.
- Comparison of intraoperative portable CT scanners in skull base and endoscopic sinus surgery: Single center case series. UT Southwestern Medical Center, 2011.
- Robot Assisted Surgery | CommonSpirit Health. CommonSpirit Health.
- Augmented reality goggles give surgeons X ray vision | New Scientist. New Scientist, 2017.
- The future of surgery: Using augmented reality goggles in the operating room. UC Davis, 2024.
- Medical Holograms Are Now Part of the Surgeon's Toolkit. Smithsonian magazine, 2015.
Selected outlets
- OpticaDesign of an Accurate Near Infrared Optical Tracking System in Surgical Navigation
- UT Southwestern Medical CenterComparison of intraoperative portable CT scanners in skull base and endoscopic sinus surgery: Single center case series
- CommonSpirit HealthRobot Assisted Surgery | CommonSpirit Health
- New ScientistAugmented reality goggles give surgeons X ray vision | New Scientist
- UC DavisThe future of surgery: Using augmented reality goggles in the operating room
- Smithsonian magazineMedical Holograms Are Now Part of the Surgeon's Toolkit
Nearby entries
- 01

X Rays in Surgery
From a shielded glass tube in Bavaria to portable scanners that refresh a surgeon's map mid operation.
- 02

Augmented Reality in Surgery
Headsets that place a patient's scans in the surgeon's view, and the regulators who now list them.
- 03
Operating at a Distance
A console in one city, a patient in another, and the fraction of a second that separates them.